Radiation Instrument Index

Classic and current radiation protection instruments, grouped by what they do. Find the meter your site uses and see the generic type the exam asks about, what it detects, its range, and the manual.

Every site has its own instrument fleet, but the DOE RCT exam asks about instrument types: a GM count rate meter with a pancake probe, an ion chamber, a ZnS alpha probe, a gas proportional PCM. Each entry below gives that generic type, so you can match what is on your shelf to what is on the test.

Specifications are taken from the manufacturer's manual or datasheet. Your site's procedures and the instrument's own calibration govern how you use it. Study more with the study guides and the practice tests.

Count Rate Meters & Contamination Probes

Friskers, pancake GM probes, alpha and beta scintillation probes. DOE RCT Module 2.17.

Eberline AC-3-7 / AC-3-8 Now Thermo Fisher Scientific

alpha scintillation probe
Detector
Alpha scintillators (CP-1 connector)
Detects
alpha

Hand probes for alpha surface contamination.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1500 V).

Eberline ASP-2 Now Thermo Fisher Scientific

microprocessor portable ratemeter for external probes
Detector
None built in; accepts GM, gas proportional and scintillation probes (MHV, BNC or CP-1 connectors)
Detects
depends on probe (alpha, beta, gamma, neutron)
Range
Probe HV 300 to 2500 V; thresholds 0.5 to 60 mV; units CPS, CPM, DPS, DPM, Bq, R/hr, Sv/hr, Gy/hr, Rem/hr

General-purpose portable ratemeter that reads out contamination or dose rate from the probe plugged into it.

Basic model: ratemeter mode only. Set up with the WinASP2 host program; 3 C cells, about 100 h.

Eberline ASP-2e Now Thermo Fisher Scientific

enhanced microprocessor portable ratemeter for external probes
Detector
None built in; accepts GM, gas proportional and scintillation probes (MHV, BNC or CP-1 connectors)
Detects
depends on probe (alpha, beta, gamma, neutron)
Range
Probe HV 300 to 2500 V; thresholds 0.5 to 60 mV; units CPS, CPM, DPS, DPM, Bq, R/hr, Sv/hr, Gy/hr, Rem/hr

General-purpose portable ratemeter that reads out contamination or dose rate from the probe plugged into it.

Adds integrated dose, scaler counts and a pulse height window for energy-proportional probes (same manual as ASP-2).

Eberline HP-100 series (HP-100C, HP-100CGS) Now Thermo Fisher Scientific

gas flow proportional contamination probe
Detector
100 cm2 gas flow proportional (MHV connector)
Detects
alpha, beta

Hand probe for alpha and beta surface contamination.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1900 V).

Eberline HP-190A Now Thermo Fisher Scientific

end window GM probe
Detector
End window GM detector (BNC)
Detects
beta

Hand probe for beta surface contamination.

Probe for the ASP-2 and ASP-2e ratemeters (up to 900 V).

Eberline HP-210 Now Thermo Fisher Scientific

"pancake" Current
pancake GM probe
Detector
pancake Geiger-Mueller detector; thin mica window 1.4 to 2.0 mg/cm2 protected by a stainless steel screen; power supply 900 +/- 50 V
Detects
alpha and beta

Hand probe for daily contamination surveys of personnel, tables, floors and equipment.

Useful beta sensitivity down to 40 keV, alpha sensitive above 3 MeV; HP-210T adds tungsten shielding for beta in gamma fields.

Eberline HP-220A Now Thermo Fisher Scientific

shielded/directional GM probe
Detector
Shielded/directional GM detector (BNC)
Detects
gamma

Shielded, directional gamma probe.

Probe for the ASP-2 and ASP-2e ratemeters (up to 600 V).

Eberline HP-260 Now Thermo Fisher Scientific

"pancake"
pancake GM probe
Detector
pancake GM tube with thin mica window 1.4 to 2.0 mg/cm2, screen 79% open; active area 15.5 cm2; operating voltage 900 +/- 50 Vdc
Detects
alpha, beta, gamma

Hand probe for contamination surveys; background sensitivity about 3600 cpm per mR/hr (Cs-137).

Same detector specification as the HP-360, which differs only in its housing.

Eberline HP-330 Now Thermo Fisher Scientific

sealed gas proportional probe
Detector
Sealed gas proportional detector (MHV connector)
Detects
alpha, beta

Hand probe for alpha and beta surface contamination without a gas bottle.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1800 V).

Eberline HP-340 Now Thermo Fisher Scientific

alpha/beta scintillation probe
Detector
Alpha/Beta scintillator (MHV connector)
Detects
alpha, beta

Hand probe for alpha and beta surface contamination.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1000 V).

Eberline HP-350 Now Thermo Fisher Scientific

alpha scintillation probe
Detector
Alpha scintillator (MHV connector)
Detects
alpha

Hand probe for alpha surface contamination.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1500 V).

Eberline HP-360 Now Thermo Fisher Scientific

pancake GM probe
Detector
Pancake GM detector (BNC)
Detects
beta

Pancake GM hand probe for contamination surveys.

Probe for the ASP-2 and ASP-2e ratemeters (up to 900 V).

Eberline HP-370 Now Thermo Fisher Scientific

pancake low energy gamma probe
Detector
Pancake low energy gamma detector (MHV connector)
Detects
low energy gamma

Hand probe for low-energy gamma emitters.

Probe for the ASP-2 and ASP-2e ratemeters.

Eberline HP-380A Now Thermo Fisher Scientific

alpha scintillation probe
Detector
100 cm2 alpha scintillator (MHV connector)
Detects
alpha

100 cm2 hand probe for alpha surface contamination.

Probe for the ASP-2 and ASP-2e ratemeters.

Eberline HP-380AB Now Thermo Fisher Scientific

alpha/beta scintillation probe
Detector
100 cm2 alpha/beta scintillator (MHV connector)
Detects
alpha, beta

100 cm2 hand probe for alpha and beta surface contamination.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1000 V).

Eberline LEG-1 Now Thermo Fisher Scientific

low energy gamma scintillation probe
Detector
Low energy gamma scintillator, 1 in. dia x 0.04 in. thick NaI crystal (CP-1 connector)
Detects
low energy gamma

Probe for low-energy gamma and X-rays.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1500 V).

Eberline PG-2 Now Thermo Fisher Scientific

low energy gamma/X-ray scintillation probe
Detector
Low energy gamma/X-ray scintillator, 2 in. dia x 0.08 in. thick NaI crystal (CP-1 connector)
Detects
gamma, X-ray (low energy)

Probe for low-energy gamma and X-ray emitters.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1300 V).

Eberline RM-14 Now Thermo Fisher Scientific

GM count rate meter
Detector
None built in; BNC detector input with 900 V high voltage, made primarily for a Geiger-Mueller (G-M) detector (scintillation detectors with modification)
Detects
as the attached detector; primarily G-M
Range
0-500 cpm meter scale with X1, X10, X100 multipliers (500; 5000; 50,000 cpm full scale)

Read out a G-M probe as a count rate with audible clicks and an adjustable locking alarm.

AC line or internal 6 V gel battery (about 50 hours); fast/slow response switch; recorder and scaler outputs.

Eberline RM-20 Now Thermo Fisher Scientific

alarming count rate meter (ratemeter)
Detector
None built in; charge-sensitive input and variable high voltage (0 to 2.5 kV scale) for Geiger-Mueller, scintillation and proportional detectors
Detects
as the attached detector (GM, scintillation, proportional)
Range
RM-20-1: 0-500, 0-5k, 0-50k, 0-500k cpm; RM-20-5: 0-10, 0-100, 0-1k, 0-10k cps

Read out a contamination probe as a count rate with audible clicks and an adjustable locking alarm.

RM-20 series units are identical except meter face; AC powered, optional internal 12 V battery; RM-25 later replaced it.

Eberline RM-25 Now Thermo Fisher Scientific

microcomputer-based count rate meter and frisker station
Detector
None built in; accepts Eberline GM, gas proportional and scintillation detectors, 500 to 2500 VDC, MHV connector
Detects
as the attached detector (GM, gas proportional, scintillator/PM tube)
Range
Analog meter marked 0 to 500 linearly; range multipliers x1, x10, x100, x1K

Frisker station: reads a contamination probe as a count rate with audible counts and alarm.

Enhanced RM-14/RM-20 replacement; setup and calibration by PC (WinRM25); 110-130 VAC or 12-14 VDC.

Eberline SPA-1A Now Thermo Fisher Scientific

alpha scintillation probe for swipes
Detector
Alpha scintillator, 1 in. diameter zinc sulfide element (MHV connector)
Detects
alpha

Alpha scintillator for particulate filters and swipe samples.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1400 V).

Eberline SPA-3 Now Thermo Fisher Scientific

gamma scintillation probe
Detector
High sensitivity gamma scintillator, 2 in. dia x 2 in. NaI crystal (CP-1 connector; SPA-3MHV has MHV)
Detects
gamma

High sensitivity gamma scintillation probe.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1600 V).

Eberline SPA-6 Now Thermo Fisher Scientific

gamma scintillation probe
Detector
Medium sensitivity gamma scintillator, 2 in. dia x 2.4 in. plastic element (MHV connector)
Detects
gamma

Medium sensitivity gamma probe.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1600 V).

Eberline SPA-8 Now Thermo Fisher Scientific

gamma scintillation probe
Detector
Gamma scintillator, 1 in. dia x 1 in. NaI crystal (MHV connector)
Detects
gamma

General gamma scintillation probe.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1500 V).

Eberline SPA-9 Now Thermo Fisher Scientific

iodine gamma scintillation probe
Detector
Gamma scintillator for iodine, 2 in. dia x 0.5 in. NaI crystal (MHV connector)
Detects
gamma (iodine)

Gamma probe for radioiodine.

Probe for the ASP-2 and ASP-2e ratemeters (up to 1600 V).

Ludlum Model 3

Current
analog GM/scintillation count rate meter (ratemeter)
Detector
no built-in detector; used with external GM or scintillation detectors; high voltage adjustable 400 to 1500 Vdc
Detects
alpha, beta, gamma, depending on the probe attached
Range
typically 0-200 mR/hr, or 0-500,000 cpm (4 linear ranges, x0.1, x1, x10, x100)

General purpose analog survey meter; the probe plugged in decides what it reads.

Fast (4 s) or slow (22 s) meter response; two D cells, typically more than 2000 hours.

Ludlum Model 12

Current
analog count rate meter for GM, scintillation or proportional detectors
Detector
no built-in detector; external GM, scintillation or proportional detectors; high voltage adjustable 400 to 2500 volts
Detects
alpha, beta, gamma, depending on the probe attached
Range
typically 0-500,000 counts/minute (CPM); multipliers x1, x10, x100, x1000

Analog ratemeter for contamination and gamma surveys; wide high voltage range lets it run gas proportional probes like the 43-68.

Input sensitivity adjustable -1 to -100 mV; fast (4 s) or slow (22 s) response.

Ludlum Model 26

"frisker" Current
integrated GM pancake frisker (meter and probe in one housing)
Detector
pancake GM (Geiger-Mueller) detector, stainless steel screen (79% open); active window area 15.51 cm2, open 12.26 cm2
Detects
alpha, beta, gamma; energy range 16.3 keV to 1.3 MeV
Range
0.1 cps to 1.99 kcps; 1 cpm to 99.9 kcpm

One-handed frisking of people and objects for contamination.

RATE, MAX and COUNT modes (count up to 20 min); two AA batteries, about 1000 hours; Model 26-3 adds mR/hr and dpm units.

Ludlum Model 43-5

Current
ZnS alpha scintillation probe
Detector
ZnS(Ag) scintillator, window typically 0.8 mg/cm2 metalized polyester; active area 79 cm2 (open 52 cm2); 38 mm photomultiplier tube
Detects
alpha

Alpha-only survey probe for surfaces; efficiency 13% (4pi) for Pu-239, background 3 cpm or less.

Operating voltage typically 500 to 1200 V; used with general purpose ratemeters and scalers.

Ludlum Model 43-68

"gas proportional probe" Current
gas proportional alpha-beta probe
Detector
rechargeable gas proportional detector, P-10 counting gas; window typically 0.8 mg/cm2 metalized polyester; active area 126 cm2 (open 100 cm2)
Detects
alpha and beta (less than 1% gamma)

Large-area alpha/beta contamination probe run on P-10 gas with a ratemeter or the Model 2360.

4pi efficiency Pu-239 20%, Sr-90/Y-90 20%; alpha plateau typically 1100-1400 V, beta-gamma 1600-1800 V.

Ludlum Model 43-89

Current
ZnS/plastic alpha-beta scintillation probe
Detector
ZnS(Ag) adhered to 0.25 mm plastic scintillator; window 1.2 mg/cm2 metallized polyester; active area approx. 125 cm2 (open approx. 100 cm2)
Detects
simultaneous alpha and beta

Large-area alpha and beta contamination probe for floors, tools and equipment.

4pi efficiency Pu-239 17.5%, Tc-99 15%, Sr-90/Y-90 17.5%; suggested instruments Model 2360, 3002, 3003.

Ludlum Model 43-93

Current
ZnS/plastic alpha-beta scintillation probe
Detector
ZnS(Ag) adhered to 0.25 mm plastic scintillator; window 1.2 mg/cm2 metalized polyester; active area 100 cm2 (open 88 cm2)
Detects
simultaneous alpha and beta

Alpha and beta contamination probe read separately on a dual-channel meter such as the 2360.

4pi efficiency Tc-99 15%, Sr-90/Y-90 20%, Pu-239 20%; background alpha 3 cpm or less, beta 300 cpm or less.

Ludlum Model 44-2

Current
NaI gamma scintillation probe
Detector
scintillator, 2.5 x 2.5 cm (1 x 1 in.) (diameter x thickness); 2.9 cm photomultiplier tube
Detects
gamma; recommended energy range 20 keV to 1.5 MeV

Sensitive gamma probe for finding sources and low-level gamma surveys; about 175 cpm per µR/hr (Cs-137).

Background 1800 cpm in a 10 µR/hr field; energy dependent, so not for dose measurements without correction.

Ludlum Model 44-9

"pancake" Current
pancake GM probe
Detector
pancake-type, halogen-quenched GM; mica window 2.0 +/- 0.3 mg/cm2; active window area 15.51 cm2 (open 12.26 cm2)
Detects
alpha, beta, gamma

Standard hand probe for frisking and contamination surveys; 3300 cpm per mR/hr (Cs-137 gamma).

Typical 4pi efficiency 22% Sr-90/Y-90, 5% C-14; operates at 900 Vdc; optional dose filter flattens energy response.

Ludlum Model 2241-2

Current
digital survey meter with built-in scaler (ratemeter/scaler)
Detector
no built-in detector; stores settings for two external detectors (DET1/DET2 switch)
Detects
alpha, beta, gamma, depending on the probe attached
Range
auto-ranging 0.0 µR/hr to 9999 R/hr; 0.000 µSv/h to 9999 Sv/h; 0 C/m to 999 C/m; 0 C/s to 100 kC/s

Digital ratemeter and scaler; swap between a pancake GM and a gamma scintillator without recalibrating.

Ratemeter or scaler mode (1 to 9999 s); two D cells, typically 200 hours; 2241-2RK kit adds a 44-9 and 44-2.

Ludlum Model 2360

Current
dual-channel alpha/beta scaler/ratemeter data logger
Detector
no built-in detector; suggested detectors are dual phosphor scintillation and gas proportional probes
Detects
simultaneous alpha and beta
Range
meter dial 0-500 cpm; range multipliers x1, x10, x100, x1000; 6-digit scaler, count times 0.1 to 60 minutes

Counts alpha and beta together from one probe, with logged results for contamination surveys and smear counting.

Stores 550 data points; RS-232 port; six separate alarms; typical battery life 250 hours.

Mirion RDS-80

Current
GM surface contamination meter
Detector
end window GM tube, 1.5 - 2 mg/cm2 mica window, active surface area 15.2 cm2
Detects
alpha >2 MeV, beta Emax >100 keV, gamma and X-rays 5 keV to 3 MeV
Range
surface contamination 1 to 100 000 cps or 0.01 to 1 000 000 Bq/cm2

Handheld pancake-style contamination meter that reads cps or Bq/cm2.

Histogram of 480 points; IrDA port with RDS-CSW software; 2 AA batteries, at least 2000 hours.

Thermo Scientific RadEye B20

Current
pancake GM multi-purpose survey meter (contamination and dose rate)
Detector
1 pancake GM tube, window diameter 44 mm (1.7 in.), 1.8 - 2.0 mg/cm2
Detects
alpha, beta, gamma and X-ray
Range
contamination 0 - 10 kcps (B20), 0 - 500 kcps (B20-ER); dose rate 0.2 µSv/h - 2 mSv/h (B20), 0.2 µSv/h - 100 mSv/h (B20-ER)

Pocket frisker and survey meter; scaler, ratemeter and dose rate modes.

Energy range 17 keV - 3 MeV with optional filters; optional alpha blocker separates alpha from beta; 300 g, 2 AAA batteries.

Dose Rate Meters

Ion chambers, energy compensated GM meters, micro-R meters. DOE RCT Module 2.16.

Bicron Micro-Analyst

"micro R meter"
micro-R scintillation survey meter with single channel analyzer
Detector
Internally mounted 1 x 1 in. NaI scintillator
Detects
gamma, X-ray
Range
0 - 5000 µR/h in four linear ranges (x1, x10, x100, x1000)

Low-level gamma survey in µR/h; SCA mode reads only a chosen energy window to cut background.

Accuracy within 10% of reading (20 to 100% of full scale); reads µR/h based on Cs-137; 9 V battery.

Bicron Micro Rem Now Thermo Fisher Scientific

Current
tissue-equivalent scintillation dose rate meter
Detector
Internally mounted tissue-equivalent organic scintillator (sensitive area extends outside the case with the Extended Detector option)
Detects
gamma and X-ray, 40 keV to 1.3 MeV (17 keV to 1.3 MeV with Low Energy option)
Range
Meter scale 0-200 µrem/h; five ranges X0.1 0-20, X1 0-200, X10 0-2000, X100 0-20,000, X1000 0-200,000 µrem/h (Micro Sievert model 0-0.2 to 0-2000 µSv/h)

Read low-level gamma dose rate directly in rem, with no mR/h conversion.

Micro Sievert model reads µSv/h; eight part numbers (standard or extended probe, standard or low energy); two 9 V batteries.

Bicron RSO-5

ion chamber dose rate meter
Detector
Air-filled ionization chamber vented to atmosphere through silica gel desiccant, approx. 200 cm3, 200 mg/cm2 phenolic; 7 mg/cm2 mylar window; sliding beta shield 400 mg/cm2 phenolic
Detects
beta, gamma and X-ray
Range
0-5000 mR/h in 4 linear ranges (5, 50, 500, 5000 mR/h)

Survey gamma and X-ray exposure rate, and beta with the shield open, in plant areas.

Harshaw/Bicron, Saint-Gobain; one 9 V battery; collection potential self-test; energy response +/-20%, 12 keV to 7 MeV.

Bicron RSO-50

ion chamber dose rate meter
Detector
Air-filled ionization chamber vented to atmosphere through silica gel desiccant, approx. 200 cm3, 200 mg/cm2 phenolic; 7 mg/cm2 mylar window; sliding beta shield 400 mg/cm2 phenolic
Detects
beta, gamma and X-ray
Range
0-50 R/h in 4 linear ranges (50 R/h, 5 R/h, 500 mR/h, 50 mR/h)

Survey higher gamma and X-ray exposure rates than the RSO-5 reaches.

High-range counterpart of the RSO-5; one 9 V battery; collection potential 30 V from ten energy cells.

Eberline HP-270 Now Thermo Fisher Scientific

energy compensated GM probe
Detector
Energy compensated GM detector (BNC)
Detects
gamma
Range
0 - 3 R/Hr

Probe for gamma exposure rate measurement.

Probe for the ASP-2 and ASP-2e ratemeters (up to 900 V).

Eberline HP-290 Now Thermo Fisher Scientific

energy compensated GM probe
Detector
Energy compensated GM detector (BNC)
Detects
gamma
Range
0 - 80 R/Hr

Probe for higher gamma exposure rates.

Probe for the ASP-2 and ASP-2e ratemeters (up to 550 V).

Eberline RO-2 Now Thermo Fisher Scientific

air ionization chamber dose rate meter
Detector
Air-filled ionization chamber vented to atmospheric pressure, 3 in. (7.62 cm) diameter, 208 cm3; 200 mg/cm2 phenolic wall inside 0.05 in. aluminum case; 7 mg/cm2 Mylar window; sliding beta shield 400 mg/cm2
Detects
beta, gamma, X-ray
Range
0-5, 0-50, 0-500, 0-5000 mR/h (four linear ranges)

Survey gamma and X-ray exposure rate and beta dose rate in plant areas; calibrated to Cs-137 gamma.

Slide shield open for beta; three 9 V batteries; energy response +/-20%, 12 keV to 7 MeV.

Eberline RO-2A Now Thermo Fisher Scientific

air ionization chamber dose rate meter
Detector
Air-filled ionization chamber vented to atmospheric pressure, 3 in. (7.6 cm) diameter, 208 cm3; 200 mg/cm2 phenolic walls inside 0.05 in. aluminum case; 7 mg/cm2 Mylar window; sliding beta shield 400 mg/cm2
Detects
beta, gamma, X-ray
Range
0-50, 0-500 mR/h, 0-5, 0-50 R/h (four linear ranges, 50 mR/h to 50 R/h full scale)

Survey higher gamma and X-ray exposure rates and beta dose rate than the RO-2 reaches.

Higher-range version of the RO-2; four 9 V batteries.

Eberline RO-20 Now Thermo Fisher Scientific

Legacy
air ionization chamber dose rate meter
Detector
Air-filled ionization chamber vented to atmospheric pressure, 2.84 in. (7.21 cm) inside diameter, 13.4 in3 (220 cm3); conductive plastic wall, about 1000 mg/cm2 total; two-layer Mylar window about 7 mg/cm2; sliding beta shield about 1000 mg/cm2
Detects
beta, gamma, X-ray
Range
0-5, 0-50, 0-500 mR/h and 0-5, 0-50 R/h (five linear ranges)

Survey X-ray and gamma exposure rate (and beta with slide open) in plant areas; calibrated to Cs-137 gamma.

Five ranges (the RO-2 has four); back-lit meter; five C cells plus 30 V chamber bias cells. Discontinued by Thermo; the RO-20 AA is the AA-cell version.

Eberline RO-20 AA Now Thermo Fisher Scientific

air ion chamber survey meter
Detector
Air ionization chamber, 220 cm3 (7.21 cm diameter), vented to atmosphere; about 7 mg/cm2 Mylar window with sliding beta shield
Detects
beta, gamma, X-ray
Range
Five linear ranges: 0-5, 0-50, 0-500 mR/h and 0-5, 0-50 R/h

Exposure rate survey from low mR/h to 50 R/h; open the beta shield to read beta dose rate.

AA-cell version of the RO-20; calibrated to Cs-137; reads about 8% of true neutron field (PuBe); 5 s response.

Ludlum Model 9-3

Current
air ionization chamber dose rate meter
Detector
air ionization chamber, 220 cm3 (13.4 in3); window 7 mg/cm2 metalized polyester; retractable 1000 mg/cm2 beta shield
Detects
beta and gamma (gamma to 6 keV with slide open)
Range
typically 0.2 - 50000 mR/hr (5 ranges x1 to x10k)

Beta-gamma dose rate meter for higher fields; slide open reads beta plus gamma, slide closed reads gamma only.

Energy response within 20% from 40 keV to 2 MeV; two AA batteries.

Ludlum Model 14C

Current
analog ratemeter with internal high-range GM
Detector
internal energy compensated GM, used on x1000 scale only; external GM or scintillation detector for x0.1 to x100 scales
Detects
alpha, beta, gamma (probe dependent); internal tube reads gamma
Range
internal GM 0-20 mSv/h (0-2000 mR/hr); external detector ranges x0.1, x1, x10, x100

Survey meter that also has a built-in high-range gamma tube so the needle pins in very high fields.

Internal detector energy response within 15% from 60 keV to 3 MeV; set for 900 V GM operation.

Ludlum Model 19

"micro R" Current
scintillation microR meter
Detector
internal 2.5 x 2.5 cm (1 x 1 in.) NaI(Tl) scintillator
Detects
gamma (low-level)
Range
0 - 5000 µR/hr (ranges 0-25, 0-50, 0-250, 0-500, 0-5000 µR/hr)

Finds and measures near-background gamma levels; sensitivity about 175 cpm per µR/hr (Cs-137).

Energy response is nonlinear, so dose and exposure readings can be over or under reported.

Mirion RAM ION

portable ion chamber survey meter
Detector
ion chamber
Detects
gamma, X-ray and beta
Range
dose rate 0.1 mR/h to 50 R/h (1 µSv/h - 500 mSv/h); integrated dose 1 µR to 1000 R (0.01 µSv - 10 Sv)

Ion chamber dose rate and dose meter with analog bar graph and alarms.

Mirion RDS-31

Current
multi-purpose energy compensated GM survey meter
Detector
one energy compensated GM tube, response according to ambient dose equivalent H*(10); external alpha, beta and gamma probes can be connected
Detects
gamma and X-rays, 48 keV to 3 MeV; alpha and beta with external probes
Range
dose rate 0.01 µSv/h to 0.1 Sv/h (1 µrem/h to 10 rem/h); dose 0.01 µSv to 10 Sv

Handheld H*(10) dose rate meter that doubles as a contamination meter with probes.

IP67; complies with ANSI N42.33 and IEC 60846; 2 AA batteries; 220 g with batteries.

Thermo Scientific FH 40 G

Current
multi-purpose digital survey meter
Detector
internal proportional counter tube for gamma with optional external plug-and-play probes
Detects
gamma, X-ray, beta (with probe)
Range
FH 40 G: 1 µR/h to 100 R/h, 36 keV to 1.3 MeV; FH 40 G-10: 10 nSv/h to 1 Sv/h, 30 keV to 4.4 MeV

General gamma dose rate meter; base unit for neutron (WENDI-2), gamma and contamination probes.

IP 67; 2 AA batteries; datalogging of 256 points; versions G, G-L, G-10, G-L10 and G-X (probe only).

Thermo Scientific RadEye G

Current
energy compensated GM dose rate meter
Detector
energy compensated Geiger-Mueller tube
Detects
gamma
Range
50 µR/h to 10 R/h (G with Black Label); energy range 45 keV to 1.3 MeV (+/-30%)

Pocket dose rate meter for gamma surveys; G-10 versions read ambient dose equivalent H*(10) in Sv.

Sold in several versions (G, G-10, GF, Ex); ranges differ by version.

Thermo Scientific RadEye PRD

Current
personal radiation detector (pager)
Detector
CsI(Tl); the PRD-ER4 adds a high dose rate detector
Detects
gamma and X-ray; neutron indication via prompt gamma analysis (PRD4 page)
Range
10 nSv/hr to 10 Sv/hr (1 µR/hr to 1000 R/hr); energy 58 keV to 6 MeV for the search detector

Worn or hand-held pager for finding sources and warning of dose rate.

Current generation is the PRD4 / PRD-ER4. The original PRD read up to 25 mR/h and the PRD-ER up to 10 R/h.

Victoreen 451B Now Fluke Biomedical

Current
ion chamber survey meter with beta slide
Detector
349 cc air ionization chamber; phenolic wall 246 mg/cm2; Mylar window 6.6 mg/cm2; beta slide 440 mg/cm2
Detects
alpha above 7.5 MeV, beta above 100 keV, gamma above 7 keV
Range
0 mR/h to 5 mR/h; 0 to 50 mR/h; 0 to 500 mR/h; 0 R/h to 5 R/h; 0 R/h to 50 R/h (auto-ranging)

Beta-gamma dose rate: slide open for skin dose, slide closed for deep dose.

Not pressurized; Freeze mode holds the peak rate; two 9 V batteries, 200 hours.

Victoreen 451P Now Fluke Biomedical

Current
pressurized ion chamber survey meter
Detector
230 cc pressurized ionization chamber (8 atmospheres or 125 psi)
Detects
gamma above 25 keV, beta above 1 MeV
Range
0 µR/h to 500 µR/h; 0 to 5 mR/h; 0 to 50 mR/h; 0 to 500 mR/h; 0 to 5 R/h (auto-ranging)

Low-level (µR) dose rate and dose from X-ray and gamma; used for area surveys and release checks.

Pressurized chamber ships as DOT Dangerous Goods under special permit DOT-SP 13187; two 9 V batteries, 200 hours.

High Range & Telescoping

Extendable and remote detectors for hot spots and high fields. DOE RCT Module 2.16.

Automess Teletector 6150AD-T

"teletector"
telescopic dual GM high range dose rate meter
Detector
two GM counting tubes at the telescope end: low range ZP1400 beta-gamma end window tube, high range ZP1300 (energy compensated)
Detects
photon (gamma and X-radiation); low range tube also detects beta
Range
useful range 2 µSv/h to 9.99 Sv/h; indicated analogue 0.1 µSv/h to 10 Sv/h, digital 0.00 µSv/h to 9.99 Sv/h

Reach into high fields from distance; telescope extends to about 4 m; high range tube takes over above 10 mSv/h.

Probe for the Automess 6150AD dose rate meter; the /H and /E versions read H*(10).

MGP Instruments TelePole WR (Wide Range) Now Mirion Technologies

"telepole"
telescoping pole GM gamma survey meter
Detector
Two GM tubes: ZP-1301 (high range) and ZP-1201 (low range), switched automatically
Detects
gamma
Range
0.05 mR/h to 1000 R/h (0.5 µSv/h to 10 Sv/h)

Reads high gamma fields from a distance on an extendable pole, for hot spots and high-dose areas.

Pole extends to 11 ft (337 cm); tubes switch at 1,500 mR/h rising.

Mirion Telepole II

Current
telescoping GM gamma survey meter
Detector
internal energy compensated GM tube (ZP-1201 or equivalent); interchangeable smart external detectors (very high range, wide range, XDS, gamma-beta)
Detects
gamma and X-rays (beta with the gamma and beta wide range detector)
Range
0.05 mR/h to 1000 R/h overall; internal GM 0.1 µSv/h to 10 mSv/h (10 µR/h to 1.5 R/h)

Survey high-dose-rate or out-of-reach areas from the end of an 11 ft pole.

Pole 120 cm collapsed, 335 cm extended; LED on the head; WRM2 radio; four AA batteries, 85 hours.

Neutron Survey Meters

Rem balls, cylinders and wide-energy neutron meters. DOE RCT Module 2.16.

Eberline HP-280 Now Thermo Fisher Scientific

neutron sphere probe
Detector
Neutron sphere, 3 inch diameter (MHV connector)
Detects
neutron

3 in. moderator sphere probe for neutron dose rate.

Probe for the ASP-2 and ASP-2e ratemeters (up to 2300 V).

Eberline NRD Now Thermo Fisher Scientific

"rem ball" Current
9 in. moderated neutron rem ball (He-3 or BF3 tube)
Detector
moderated He-3 tube in a 9 in. rem ball (BF3 tube version also existed)
Detects
neutron; 0.025 eV to 10 MeV
Range
1 to 10,000 mrem/h

Classic 9 in. rem ball for neutron dose rate, read on a survey meter such as the ASP-2.

The BF3 version (NRD9) is discontinued; the He-3 version is still sold.

Eberline NRD-1 Now Thermo Fisher Scientific

neutron sphere probe
Detector
Neutron sphere, 9 inch diameter (MHV connector)
Detects
neutron

9 in. moderator sphere probe for neutron dose rate.

Probe for the ASP-2 and ASP-2e ratemeters (up to 2300 V).

Ludlum Model 12-4

"rem ball" Current
moderated He-3 neutron dose ratemeter
Detector
He-3 proportional detector, 1.6 x 2.5 cm (0.6 x 1.0 in.), inside a 22.9 cm (9.0 in.) diameter cadmium loaded polyethylene sphere
Detects
neutron; thermal through 7 MeV, response up to 12 MeV
Range
0-100 mSv/h (0-10,000 mrem/hr)

Standard rem ball for neutron dose rate around reactors, sources and casks.

Sensitivity typically 10 cpm per µSv/h (bare AmBe); gamma rejection under 10 cpm through 10 R/hr; 8.3 kg.

Ludlum Model 42-5

"Bonner sphere" Current
neutron detector and ball cart (Bonner sphere set)
Detector
4 mm x 4 mm LiI crystal on a 3.8 cm photomultiplier tube, with 2, 3, 5, 8, 10 and 12 in. polyethylene moderator spheres
Detects
neutron; continuous response thermal to approximately 12 MeV

Neutron fluence and dose spectrum work with interchangeable moderator spheres on a cart.

Ludlum lists 42-5 as a ball cart with detector, not one fixed rem ball; about 45 cpm per mrem/hr (AmBe, 10 in.).

Ludlum Model 42-41L

"PRESCILA" Current
proton recoil scintillator neutron detector
Detector
PRESCILA proton recoil scintillator; 2.9 cm photomultiplier tube
Detects
neutron; thermal to 100 MeV

Light hand-held alternative to a rem ball for neutron dose surveys, used with a Ludlum ratemeter such as the 2363.

The 42-41 is the same detector family. About 350 cpm per mrem/hr (AmBe); 2.3 kg.

Nuclear Research Corporation (NRC) NP-2

"Snoopy"
rem-reading neutron survey meter (moderator with BF3 counter)
Detector
BF3 proportional counter, 2 in. active length, fill pressure 60 cm Hg, in a polyethylene moderator with boron-loaded attenuator
Detects
neutron, thermal (0.025 eV) to 15 MeV; no response to gamma in fields up to 500 R/h of Cs-137
Range
0-2, 0-20, 0-200, 0-2000 mrem/hr (four linear ranges)

Measure neutron dose equivalent rate in mrem/hr near reactors, accelerators and neutron sources.

Military designation AN/PDR-70; manual reissued by Canberra Industries (2003); three D cells; 25 lbs; accuracy +/-10%.

Thermo Scientific FHT 762 Wendi-2

"WENDI-2" Current
wide energy neutron dose rate detector (moderated He-3)
Detector
large He-3 tube in a WENDI-2 design moderator (design originally from Los Alamos)
Detects
neutron; energy range 25 meV to 5 GeV per ICRP 74
Range
0.01 µSv/h to 100 mSv/h (Cf-252)

Neutron dose rate across thermal to high energy, plugged into an FH 40 G or FHT 6020 monitor.

Sensitivity 0.84 cps/(µSv/h) Cf-252; weight 13.5 kg; angular dependence +/-20%.

Personnel, Portal & Tool Monitors

PCMs, portal monitors, gamma exit monitors, small article monitors. DOE RCT Module 2.17.

Eberline PCM-1C Now Thermo Fisher Scientific

"PCM"
gas flow proportional personnel contamination monitor
Detector
fifteen (15) independent gas flow proportional detectors, P-10 counting gas
Detects
beta-gamma; alpha optional (3 or 15 alpha channels)

Whole body contamination check at an RCA exit: the person counts one side, turns, and counts the other.

The PCM-1B is the same family; the manual's drawings cover both. A person is counted twice, once per side.

Eberline PCM-2 Now Thermo Fisher Scientific

whole-body personnel contamination monitor (gas proportional)
Detector
16 gas proportional detectors; 34 counting zones, each with separate alpha and beta/gamma channels; up to 75 sum zones
Detects
alpha, beta/gamma

Step-in monitor that checks the whole body for alpha and beta/gamma contamination and shows where on a body outline.

Optional hand-held frisking probe; two count cycles per person.

Ludlum Model 52

"portable portal monitor" Current
portal personnel contamination monitor (beta/gamma)
Detector
18 Geiger-Mueller detectors: 14 G-M pancake detectors in the frame (head and body) and 4 cylindrical G-M detectors in the base (feet)
Detects
beta, gamma (personnel contamination)
Performance
Count time adjustable 1 to 60 seconds (default 6 s); counting capacity 9999 counts per minute

Walk-in beta/gamma contamination check of a person at an exit or emergency response site.

Assembles in 5 minutes without tools; factory preset detects 1.0 µCi Cs-137 beta source in 10 µR/hr background; 120 VAC or six D cells.

Mirion Argos-5AB

gas flow whole body contamination monitor (two-step)
Detector
25 gas flow proportional detectors (LFP-579), window multilayer aluminized Mylar at typically 0.8 mg/cm2
Detects
alpha, beta (optional gamma Zeus option, 3 plastic scintillators)

Walk-in whole body frisk of alpha/beta contamination when leaving a controlled area.

Argos-3AB is the 18 detector version, field upgradeable to 5AB. Uses P-10 style counting gas.

Mirion Argos-5PAB

Current
gasless whole body contamination monitor (two-step)
Detector
25 TPS-AB-579 thin plastic scintillator detectors, window multilayer aluminized Mylar at 1.2 mg/cm2
Detects
alpha, beta (optional gamma Zeus option, 3 plastic scintillators)

Walk-in whole body alpha/beta contamination frisk with no counting gas to supply.

Argos-3PAB is the 18 detector version. Alpha and beta discrimination; same form factor as gas flow Argos-AB.

Mirion Cronos-4

Current
gamma object/tool monitor (small article monitor)
Detector
six 45.7 x 45.7 x 5.1 cm plastic scintillators with PMTs, 65.1 L total detector volume; 128.7 L counting chamber; 25 mm lead on six sides (optional 50 mm)
Detects
gamma >50 keV
Performance
Time to reach MDA of 83 Bq (5000 dpm) at 80 nSv/h background, 1 in. lead: 48 s for 137Cs, 10 s for 60Co

Clears tools, bags, hard hats and other articles for release from a controlled area by gamma counting.

Built-in 100 kg scale. Single or dual door. Cronos-1 is the smaller 42.9 L unit.

Mirion Cronos-11

Current
gamma object/tool monitor (large article monitor)
Detector
four 61 x 74.9 x 5.1 cm plastic scintillators (main unit) plus door scintillators, 130.5 L total detector volume; 345.5 L counting chamber; 25 mm lead on six sides
Detects
gamma >50 keV
Performance
Time to reach MDA of 83 Bq (5000 dpm) at 80 nSv/h background, 1 in. lead: 130 s for 137Cs, 22 s for 60Co

Clears larger items such as waste bags and equipment for release by gamma counting.

Same software and electronics family as Argos, GEM-5 and Sirius. Built-in 100 kg scale.

Mirion GEM-5

Current
gamma exit monitor (portal)
Detector
8 identical large plastic scintillators (3 per side, 1 top, 1 bottom), 24 sum zones; 25 mm lead on side detectors
Detects
gamma
Performance
Point source anywhere on person at 80 nGy/h background: walkthrough 830 Bq 60Co / 1850 Bq 137Cs; pause (4 s) and count 555 Bq 60Co / 830 Bq 137Cs; two-step 370 Bq 60Co (8 s) / 370 Bq 137Cs (18 s)

Screens people leaving contaminated or potentially contaminated areas for gamma-emitting contamination.

Three modes: walkthrough, pause-and-count, and two-step with a 180 degree turn. Complies with IEC 61098.

Mirion Sirius-5

Current
hand, cuff and foot contamination monitor
Detector
six detectors, ~579 cm2 window each: LFP-579 gas flow proportional (5AB) or TPS gasless plastic scintillator (5PB/5PAB/5PBG); two moveable hand detectors
Detects
alpha/beta (AB, PAB); beta (PB); beta/gamma (PBG)

Hand, cuff and foot frisk at the exit of a contaminated area.

Gas flow window is 0.8 mg/cm2 Mylar; gasless window is 1.2 mg/cm2. Optional frisker probe.

RADOS CheckPoint:Body RTM110 Now Mirion Technologies

hand, foot and clothing contamination monitor
Detector
Either RGZ gas flow detectors (P10, P7.5 or Ar/CO2, thin Mylar foil) or RBP beta plastic scintillation detectors; hand, clothing probe and foot detectors
Detects
beta, gamma; alpha with RGZ gas detectors
Performance
MDA, RGZ gas hand detector, 5 s count: Sr-90/Y-90 10 Bq, Cl-36 15 Bq, Co-60 25 Bq, Am-241 25 Bq (5 Bq discriminated)

Walk-up monitor that counts hands, feet and clothing probe on exit from a controlled area.

Manufactured by Mirion Technologies (RADOS) GmbH. Gas version needs counting gas; warm-up about 2 h with P10.

RADOS CheckPoint:Body TwoStep-Exit Now Mirion Technologies

whole-body contamination exit monitor (two-step portal)
Detector
RADOS Fibre (RFD) detectors, BetaFibre and/or GammaFibre; head, hand/forearm, body, foot and small items detectors; 10 or 20 mm lead shielding
Detects
beta and/or gamma

Step-in body monitor that checks head, hands, body and feet for contamination before leaving a controlled area.

Split-Delta geometry housing; optional small items box and head detector. Made by Mirion Technologies (RADOS) GmbH.

RADOS CheckPoint:Waste Tool&Object-Monitor (RTM606) Now Mirion Technologies

tool and small object release monitor
Detector
RFD 1600G gamma plastic fibre detectors, 400 x 400 x 50 mm, 4 or 6 around an 80 litre measurement chamber; 25 or 50 mm lead shield
Detects
gamma

Release measurement of tools and objects leaving a controlled area, with optional weight cells and nuclide vectors.

Compact class of the CheckPoint:Waste family; one or two doors; 4 or 6 detector versions. Made by Mirion Technologies (RADOS) GmbH.

Thermo Scientific iPCM12

installed personnel contamination monitor
Detector
up to 24 identically sized detectors (21 standard) in four detection zones; almost 17,000 cm2 detection area; optional gamma kit with up to nine scintillators
Detects
alpha, beta (gamma with option)

Stand-in whole body, hand and foot contamination monitor; Thermo's successor to the PCM2 geometry.

Combines the PCM-2 detector geometry and IPM9 serviceability with Series 12 electronics. 660 lb open frame.

Thermo Scientific PM12

personnel contamination portal monitor (gamma)
Detector
Eight large plastic scintillation detectors (NE110 equivalent), 2 in. thick, 12 x 22 in. each, around a 28 x 78 in. portal opening; 0.5 or 1 in. lead
Detects
gamma; energy response from 20 keV upwards

Walk-through portal that screens people for gamma contamination, with a statistical alarm set to a reliably detectable activity.

Variants PM12A (LCD), PM12B (no LCD), PM12C (power doors), each with 05L or 10L lead. Battery backup for 6 hours.

Thermo Scientific SAM12

small articles monitor (gamma)
Detector
Four or six large plastic scintillators (BC408), 15 x 15 x 2.25 in., around a stainless steel lined chamber; 1 or 2 in. lead shielding
Detects
gamma; energy response from 20 keV upwards

Cabinet that monitors tools, clothing, personal effects and waste for gamma contamination.

SAM12A standard; SAM12L low-energy plastic liner version. Variants set by detectors (4C/6C), doors (1D/2D), lead (1L/2L).

Air Samplers

High volume, low volume and lapel samplers. DOE RCT Module 2.18.

HI-Q Environmental Products CF-900 Series (CF-901, CF-902, CF-903)

Current
portable low cost high volume air sampler
Detector
brushed motor with 2-stage centrifugal blower; manual speed control; instantaneous flow reading
Range
CF-901 2-12 CFM; CF-902 5-35 CFM; CF-903 10-50 CFM

Grab or continuous duty radioiodine and particulate sampling in the field.

9.5 lb aluminum cabinet. 115 or 230 VAC; DIGITAL variants add a flow display. CF-901 takes the 1.5 in. CF holders.

HI-Q Environmental Products HVP-4200AFC

Current
high volume air sampler with automatic flow control
Detector
2-stage brushless centrifugal blower, PLC controlled with mass or volumetric flow sensor; 8 x 10 in. filter (HVP-4204AFC takes 4 in.)
Range
10-50 SCFM (HVP-4204AFC: 2-15 SCFM)

Fixed station outdoor particulate air sampling at constant flow as filters load.

115 VAC 5.0 A, max vacuum 49 in. H2O. The 3-stage HVP-4300AFC is the higher vacuum version.

MSA (Zefon International) Escort ELF

personal air sampling pump with electronic flow control
Detector
Electronic laminar flow sensor (ELF) with digital flow readout
Range
0.5 to 3.0 LPM

Low-volume lapel or general air sampling pump that holds a set flow constant for filter or cartridge samples.

Flow held within +/-2.5%; 4.8 V NiCd/NiMH battery pack, 14 to 16 h recharge; check against a primary flow standard.

RADeCO H-809V

Current
grab air sampler (particulate and radioiodine)
Detector
two-stage turbine blower, calibrated rotameter; used with RADeCO filter and radioiodine cartridge holders
Range
H-809VI typical 1-8 CFM; H-809VII typical 1-30 CFM

Lightweight grab sampler for airborne particulates or particulates plus radioiodine.

8 x 10 x 7 in., 10.5 lb, UL listed. The 12/24 VDC H-809C is a separate model.

Staplex TFIA-2

Current
portable high volume air sampler (220-240 V)
Detector
universal motor and pump with variable orifice flowmeter calibrated 0-70 CFM; 4 in. filter holder
Range
0-70 cfm (0-2 cmm)

Portable high volume particulate sampling, indoor or outdoor.

TFIA-2 is the 220-240 V version of the 110-125 V TFIA. About 10 lb; 4 in. filter holder standard.

Continuous Air Monitors (CAMs)

Real-time particulate, iodine and gas monitors. DOE RCT Module 2.18.

Bladewerx SabreBPM

portable alpha/beta continuous air monitor with alpha spectrometry
Detector
Solid-state silicon, 300 µm depleted, 450 mm2 active area; 37 mm filter holder; 6.0 LPM diaphragm pump; 1024-channel MCA for alpha
Detects
alpha (Po-212, Po-214, Po-218 plus up to two isotopes of interest) and beta

Samples air on a filter and reports acute and chronic alpha/beta concentration and DAC, fitting radon daughter peaks to separate them.

Typical beta efficiency 15% for Cl-36; alpha 4-pi efficiency expected 22 to 30%. Battery run time 8 hours.

Eberline AMS-4 Now Thermo Fisher Scientific

beta particulate continuous air monitor
Detector
Sealed gas-proportional detector (Ar-CO2 radial/gas head, Ne-CO2 in-line head), mica window 2-3 mg/cm2, with gamma background detector
Detects
beta particulate (noble gas with optional gas head)
Performance
4-pi efficiency: Tc-99 9%, Sr-90/Y-90 18%; noble gas head Kr-85 6.4%, Xe-133 4.3%

Draws air through a filter and alarms on airborne beta particulate: slow/fast concentration, net beta rate, DAC-hr, stack release.

Radial inlet or in-line stack head. WinAMS software networks 1 to 12 AMS-4 units.

Mirion iCAM

Current
alpha beta continuous air monitor (mobile)
Detector
dual PIPS large area silicon; fixed or moving filter (FSLW, GFA or FMLB)
Detects
alpha, beta (and gamma background compensation)
Performance
Typical: alpha 1E-3 to 9E3 Bq/m3, beta 1 to 3.6E5 Bq/m3 (1 hour averaging); efficiency alpha 24%, beta 24% for 36Cl or 90Sr, 15% for 60Co; flow 15 to 60 l/min

Workplace air monitor that alarms on airborne alpha/beta particulate with radon/thoron compensation.

Alpha MCA spectroscopy for radon/thoron compensation; beta window 50 keV to 2.5 MeV. Part of the CAMSYS line.

Mirion iCAM-HD

Current
high flow rate alpha beta continuous air monitor
Detector
2 x CAM1700ASV PIPS detectors
Detects
alpha, beta (optional G64 silicon gamma detector for ambient gamma)
Performance
Airflow up to 120 liters/minute; efficiency alpha 21% (up to 5.7 MeV), beta 22% for 36Cl or 90Sr; measurement range in excess of 500 kBq combined alpha and beta on filter

Higher flow version of the iCAM for lower detection limits in a fixed location.

Same electronics and software family as the iCAM. 30 kg without pump.

Thermo Scientific Alpha 7A

Legacy
alpha continuous air monitor
Detector
alpha spectrometric detector (radial entry head for ambient air; in-line head for stack or process)
Detects
alpha

Early warning of airborne transuranic (Pu-238, Pu-239) releases with DAC and DAC-hour alarms.

Tracks up to 8 isotopes with spectra refreshed every second. ALPHA7L is the Amphenol connector variant.

Counting Room

Sample counters, gamma spectroscopy, liquid scintillation, in situ gamma. DOE RCT Module 2.19.

Canberra ISOCS (S573) Now Mirion Technologies

Current
in situ gamma spectroscopy system (germanium detector with ISOCS efficiency calibration software)
Detector
Germanium detector individually ISOCS characterized by Canberra (MCNP model validated with a NIST traceable source)
Detects
gamma
Performance
Efficiency characterized for photon energies 45 keV through 7 MeV, any location within a 500 m radius sphere around the detector

Measure gamma activity of drums, pipes, soil and room surfaces in place, calibrating efficiency by software for each geometry.

LabSOCS (S574) uses the same algorithms for laboratory counting.

Canberra iSolo (SOLO300U / SOLO300G) Now Mirion Technologies

low-background alpha/beta counting system (PIPS detector)
Detector
PIPS silicon, 2000 mm2 area, 300 µm sensitive depth; SOLO300G adds a guard detector
Detects
alpha, beta
Performance
Background SOLO300U: alpha 0.10 cpm, beta 33.0 cpm guaranteed; SOLO300G: alpha 0.08 cpm, beta 12.0 cpm guaranteed

Counts air sample filters and smears for alpha and beta, with radon/thoron compensation for long-lived alpha activity.

Guaranteed 50 mm efficiency: 30.0% Am-241 alpha, 22.0% Sr-90 beta. Filters up to 60 mm diameter.

Hidex 300 SL

automatic TDCR liquid scintillation counter
Detector
three PMT TDCR counter, minimum 70 mm internal lead shielding
Detects
beta, alpha
Performance
beta up to 2,000 keV; alpha up to 10,000 keV (LSC mode)

Counts H-3, C-14 and other low-level beta or alpha emitters in liquid samples; direct DPM by TDCR.

Racks 96 x 5 mL, 96 x 7 mL or 40 x 20 mL vials. 125 kg. Needs no external standard for TDCR efficiency.

Ludlum Model 2929

Current
dual-channel alpha/beta scaler
Detector
phoswich or proportional type detectors (typically used with Model 43-10-1 alpha/beta sample counter head)
Detects
alpha, beta
Performance
2 six-digit LED scalers, 0-999999 counts; timer 0.1 to 990 minutes; HV 200-2500 V

Scaler and timer for counting alpha and beta smear and air filter samples.

Independent RS-232 outputs for alpha and beta channels.

Ludlum Model 3030

Current
alpha/beta sample counter
Detector
ZnS(Ag) on plastic scintillator, 5.1 cm PMT, 20.3 cm2 area, 0.4 mg/cm2 aluminized Mylar window
Detects
alpha, beta (simultaneous)
Performance
4pi efficiency: alpha 32% 230Th, 39% 238U, 37% 239Pu; beta 8% 14C, 27% 99Tc, 29% 137Cs, 26% 90Sr/90Y; background alpha 3 cpm or less

Counts smears and air filters for alpha and beta at the same time.

Crosstalk alpha to beta 10% or less, beta to alpha 1% or less. Also sold as the 3030-2 and 3030P.

Mirion BEGe (Broad Energy Germanium)

Current
HPGe broad energy gamma spectroscopy detector
Detector
broad energy germanium crystal, 20 to 65 cm2 area and 20 to 30 mm thick, carbon composite window, iPA II intelligent preamplifier
Detects
gamma and X-ray
Performance
3 keV to 3 MeV

Counting room gamma spectroscopy with strong efficiency below 1 MeV for typical samples.

Compared by Mirion with 60% relative efficiency coaxial detectors; low energy resolution like a LEGe.

Mirion Series 6LB

Current
automatic low background gas flow alpha/beta counter
Detector
5.7 cm (2.25 in.) gas flow proportional detector, gasless plastic scintillator anti-coincidence guard, 50 or 100 sample changer
Detects
alpha, beta
Performance
Background: alpha 0.10 cpm, beta 0.9 cpm, gross 1.0 cpm; efficiency alpha 38% (210Po), beta 45% (90Sr/90Y)

Batch counting of air filters, swipes, water and soil samples for gross alpha/beta.

Descends from the Canberra/Tennelec LB5100 and directly replaces the Series 5 XLB.

ORTEC GEM series HPGe coaxial detector Now AMETEK

Current
P-type coaxial HPGe gamma spectroscopy detector
Detector
P-type coaxial high purity germanium; efficiencies to 150%; carbon fiber endcap and ICS cryocooling options
Detects
gamma (about 40 keV and upwards)
Performance
Relative efficiency to 150%

Gamma spectroscopy of environmental, effluent and bioassay samples in the counting room.

Typically paired with ORTEC MCA electronics and GammaVision software; LN2 or mechanical cooling.

ORTEC MPC-9604 Now AMETEK

Current
four-detector low background gas proportional alpha/beta counter
Detector
four 2.25 in. (5.7 cm) pancake gas flow proportional detectors, cosmic guard, 4 in. low background lead; aluminized window 80 ug/cm2
Detects
alpha, beta
Performance
Background shielded and guarded: warranted <0.1 cpm alpha, <0.1 cpm beta; typical <0.05 alpha, <0.7 beta. Efficiency e.g. 239Pu 35%, 90Sr/Y 55% (top position)

Counts swipes, air filters and water samples for gross alpha/beta at low detection limits.

Uses P-10 gas at 60 cc/min with failsafe gas monitoring; 2 in. planchets; 1200 lb.

Revvity Tri-Carb 5110TR

liquid scintillation counter
Detector
liquid scintillation counting with time-resolved (TR-LSC) background reduction and alpha/beta separation (dual PSA discriminator)
Detects
beta, alpha (low energy beta emitters such as 3H and 14C)
Performance
Sample conveyor holds up to 408 (20 mL vials) or 720 (4 or 7 mL vials); ULLCM E2/B factory minimum 500 for 3H and 1,400 for 14C

Counts tritium and C-14 in water, bioassay and smear samples.

Revvity is the former PerkinElmer/Packard Tri-Carb line. 238 kg, built-in Windows PC with QuantaSmart.

Dosimetry

Electronic dosimeters, readers and self-reading dosimeters. DOE RCT Module 2.04.

Arrow-Tech 742

Current
direct reading pocket dosimeter (0-200 R)
Detector
fiber electrometer in electrically conductive plastic ion chamber (described as carbon-fiber electroscope)
Detects
gamma and X-ray, 16 keV to 6 MeV
Range
0-200 R

Self-reading pocket dosimeter for high-range gamma exposure checks.

Accuracy +/-10% of true exposure; leakage under 1% of full scale in 24 h; ANSI N13.5 and N322.

Arrow-Tech W138

Current
direct reading pocket dosimeter (0-200 mR, sapphire window)
Detector
fiber electrometer in electrically conductive plastic ion chamber
Detects
gamma and X-ray, 16 keV to 6 MeV
Range
0-200 mR

Low-range self-reading pocket dosimeter (SRPD) for routine entries; recharged on a charger.

W138-S is the 0-2 mSv version. Arrow-Tech recommends yearly calibration.

Landauer InLight OSL dosimeter

Current
optically stimulated luminescence (OSL) badge
Detector
aluminum oxide (Al2O3:C) detectors read out with LED stimulation and a photomultiplier tube
Detects
x-ray, gamma, beta (neutron with optional CR-39)
Range
Linear from 10 µSv (1 mrem) to over 10 Sv (1,000 rem); energy 5 keV to 20 MeV; gamma/x-ray/beta minimal reporting 50 µSv (5 mrem)

Passive whole body or area badge for monthly or quarterly dose of record.

Nondestructive readout so a badge can be re-read; no annealing. Read by an InLight reader.

MGP Instruments DMC 2000 S Now Mirion Technologies

electronic personal dosimeter (gamma and X-ray)
Detector
Gamma detector (detector 1 for gamma above 50 keV)
Detects
X and gamma from 50 keV; Hp(10)
Range
Dose equivalent 1 µSv to 10 Sv; dose equivalent rate from background to 10 Sv/h

Worn electronic dosimeter showing dose and dose rate with pre-alarm and alarm; set up and read by LDM readers.

Configured with DosiMass software. Pass-by reads at 1.2 m; CR2450 battery about 1 year; IP67.

MGP Instruments DMC 2000 X Now Mirion Technologies

electronic personal dosimeter (gamma and low-energy X-ray)
Detector
Gamma and X-ray detector (detector 2 for gamma and X above 20 keV)
Detects
X and gamma from 20 keV; Hp(10)
Range
Dose equivalent 1 µSv to 10 Sv; dose equivalent rate from background to 10 Sv/h

Worn electronic dosimeter showing dose and dose rate with pre-alarm and alarm; set up and read by LDM readers.

Same manual as the DMC 2000 S; configured with DosiMass. Battery about 9 months; IP53.

MGP Instruments DMC 2000 XB Now Mirion Technologies

electronic personal dosimeter (gamma, X-ray and beta)
Detector
Gamma/X-ray detectors plus a beta detector
Detects
X and gamma from 20 keV; beta from 60 keV; Hp(10) and Hp(0.07)
Range
Dose equivalent 1 µSv to 10 Sv; dose equivalent rate from background to 10 Sv/h

Worn electronic dosimeter showing dose and dose rate with pre-alarm and alarm; set up and read by LDM readers.

Only DMC 2000 version reading shallow dose Hp(0.07). Same manual as DMC 2000 S. Battery about 9 months; IP53.

MGP Instruments DMC 3000 Now Mirion Technologies

electronic personal dosimeter (gamma and X-ray)
Detector
Two detectors: one for high energy gamma and X-ray above 50 keV, one for low energy X-ray down to 15 keV
Detects
X-ray and gamma, 15 keV to 7 MeV; Hp(10)
Range
Dose 1 µSv to 10 Sv; dose rate 0.1 µSv/h to 20 Sv/h

Worn electronic dosimeter with large backlit display, speaker, LEDs and vibrator alarms for dose and dose rate.

G2 protocol keeps it compatible with DMC 2000 readers (LDM 220/2000). AAA battery, about 9 months typical; IP67.

MGP Instruments LDM 210 Now Mirion Technologies

hands-free dosimeter reader (serial, mains powered)
Range
Dosimeter link 20 to 30 cm nominal (5 to 20 cm reduced range)

Connects DMC 2000 and SOR dosimeters to dosimetry software (DosiMass, Dosimed, Dosifast) for assigning, setup and readout.

RS-232 to PC; 9 V DC power supply, optional 2 AAA battery (about 10 h). Same manual as LDM 220.

MGP Instruments LDM 220 Now Mirion Technologies

hands-free dosimeter reader (USB powered)
Range
Dosimeter link 20 to 30 cm nominal (5 to 20 cm reduced range)

Connects DMC 2000 and SOR dosimeters to dosimetry software for assigning, setup and readout over USB.

Powered by the USB port (4.15 to 5.25 V). DMC 3000 in G2 mode is compatible. Same manual as LDM 210.

Thermo Scientific EPD-G

electronic personal dosimeter (gamma only, Mk2 family)
Detector
multi-detector silicon technology, beta window removed
Detects
X and gamma
Range
Dose 0 to >16 Sv (0 to >1600 rem); dose rate 0 to >4 Sv/h (0 to >400 rem/h); photon 15 keV to 10 MeV

Worn electronic dosimeter showing Hp(10) deep and Hp(0.07) shallow dose and dose rate with alarms.

Gamma-only variant of the Mk2 beta/gamma EPD; single AA battery. Thermo has replaced the EPD line with the EPD TruDose.

Thermo Scientific EPD TruDose

electronic personal dosimeter (G and BG versions)
Detector
multi-detector technology measuring gamma and beta; Bluetooth Low Energy telemetry
Detects
gamma (G); beta and gamma (BG)
Range
Hp(10) dose 1.0 µSv to 10 Sv; dose rate 0.05 µSv/h to 10 Sv/h (IEC61526); Hp(10) photon 16 keV to 1.5 MeV

Worn electronic dosimeter for real-time dose and dose rate with built-in Bluetooth.

IP67 (G) and IP65 (BG). Successor to the EPD Mk2 line.

Thermo Scientific Harshaw 6600 Plus

Current
automated TLD card reader
Detector
two heating channels with dual cooled PMTs; non-contact hot nitrogen gas heating; LiF:Mg,Ti cards
Detects
gamma, beta, neutron dosimeters (TLD)
Range
Seven decades; readout 70 TLD cards per hour (EXT-RAD 140/h, DXT-RAD 280/h)

Reads TLD cards and extremity dosimeters for dose of record.

Processes up to 200 cards per run; optional built-in 0.5 mCi 90Sr/Y irradiator; WinREMS software.

Radionuclide Identifiers

Handheld spectroscopic identifiers.

Mirion SPIR-Ace

Current
handheld radionuclide identification device (RIID)
Detector
NaI(Tl) 35 mm x 51 mm or LaBr3(Ce) 25.4 mm x 34 mm; energy compensated GM tube; optional 6LiZnS:Ag neutron detector
Detects
gamma, neutron (optional)
Range
25 keV to 3 MeV (gamma); gamma dose rate 0.001 µSv/h to 100 mSv/h (0.1 µR/hr to 10 R/hr)

Identifies radionuclides and estimates quantity with Genie software support.

Library of 80 nuclides, up to eight identified at once. ANSI N42.34 and IEC62327. 1.45 kg maximum.

ORTEC Detective X Now AMETEK

handheld HPGe radioisotope identifier (RIID)
Detector
P-type coaxial HPGe crystal 65 mm diameter x 50 mm length with Stirling cooler
Detects
gamma (optional neutron: segmented LiF/ZnS)
Range
Gamma dose rate range <0.05 mR/h to >100 R/h

High resolution identification of radionuclides in the field.

Library of over 175 nuclides; initial cool down about 8 hours; 6.98 kg (gamma only); IP65.

target systemelectronic GmbH identiFINDER NHG

handheld radionuclide identifier (NaI gamma spectrometer)
Detector
1.4 in. x 2 in. NaI(Tl) scintillator; internal GM tube for high dose rates; optional moderated He-3 tube for neutrons
Detects
gamma spectrometry and dose rate; neutron indication (optional)
Range
Dose rate 0.01 µSv/h to 1 Sv/h (NaI to 500 µSv/h, GM above); energy range 15 keV to 3 MeV

Identifies gamma nuclides from the spectrum, reads dose rate, and finds sources.

Resolution 7.0 to 7.5% at 662 keV; internal Cs-137 stabilization; 100 spectra memory.

Thermo Scientific RIIDEye X

handheld radioisotope identifier (RIID)
Detector
2 in. x 2 in. NaI (X-G, X-GN) or 1.5 in. x 1.5 in. LaBr (X-H, X-HN); optional CLYC neutron crystal
Detects
gamma, neutron (N versions)
Range
20 keV to 3 MeV

Identifies the radionuclide behind an unknown source or alarm.

2.6 kg with NaI. IP65; tested to ANSI N42.34-2006. Original RIIDEye (2 in. NaI, 1.4 kg) was the earlier model.

Area & Remote Monitoring

Area radiation monitors, wireless and teletechnician systems.

MGP Instruments AMP-50 Now Mirion Technologies

GM gamma area monitor probe with handheld display meter
Detector
Energy compensated GM tube (ZP1201 or equivalent)
Detects
gamma, X-ray
Range
0.05 mR/h to 4000 mR/h

Remote gamma rate probe on a cable (up to 350 ft) for tanks, filters and underwater areas; reads on a handheld meter or feeds a monitor/WRM.

Sensitivity 17 cps/mR/h (Cs-137); 70 keV to 2.0 MeV; detector waterproof to 20 m; part RTM-A007.

MGP Instruments AMP-100 Now Mirion Technologies

GM gamma area monitor probe with handheld display meter
Detector
Energy compensated GM tube (ZP1301 or equivalent)
Detects
gamma, X-ray
Range
0.005 R/h to 1000 R/h

Remote gamma rate probe on a cable (up to 350 ft) for tanks, filters and underwater areas; reads on a handheld meter or feeds a monitor/WRM.

Sensitivity 300 cps/R/h (Cs-137); 70 keV to 2.0 MeV; detector waterproof to 20 m; part RTM-A004.

MGP Instruments AMP-200 Now Mirion Technologies

GM gamma area monitor probe with handheld display meter
Detector
Energy compensated GM tube (4G60M or equivalent)
Detects
gamma, X-ray
Range
1.0 R/h to 10,000 R/h

Remote gamma rate probe on a cable (up to 350 ft) for tanks, filters and underwater areas; reads on a handheld meter or feeds a monitor/WRM.

Sensitivity 77 cps/R/h (Cs-137); 70 keV to 2.0 MeV; detector waterproof to 20 m; part RTM-A005.

MGP Instruments DMC 3000 Telemetry Module Now Mirion Technologies

wireless telemetry transmitter add-on for the DMC 3000
Range
Transmission range 1.6 miles line of sight (900 MHz), 1 mile (2.4 GHz DM); interval 2 to 64 s

Clips onto a DMC 3000 to send its readings to a base station (WRM2 protocol) and add bright LED alarm indicators.

Works with WinWRM2, TeleView 2000 and TeleView 3000 software; AAA battery, over 24 hours of runtime.

MGP Instruments WRM2 Wireless Remote Monitoring System Now Mirion Technologies

wireless remote monitoring (teledosimetry) system
Detects
dose data from DMC-2000 electronic dosimeters and area monitors
Range
Radio range 1.6 miles line of sight (900 MHz); 1,150 ft (350 m) at 10 mW (2.4 GHz)

Sends worker dosimeter readings and alarms by radio to a base station so HP can watch jobs remotely.

Set up with the WRM Configuration Studio software.

Calibration & Support Equipment

Flow, pressure and temperature references used to calibrate the instruments above.

Hart Scientific (Fluke) 1620A DewK

"DewK"
thermo-hygrometer (temperature and relative humidity reference)
Detector
Two detachable, interchangeable sensors, each measuring temperature and relative humidity
Range
Temperature 0 C to 50 C; relative humidity 0% to 100% RH

Measures and logs temperature and humidity at two points, with alarms, to document environmental conditions.

H model calibrated +/-0.125 C and +/-1.5% RH; S model +/-0.25 C and +/-2% RH. Stores about 400,000 readings.

Heise PTE-1 Now Ashcroft Inc.

handheld pressure calibrator
Detector
Two bays for Heise HQS Quick-Select modules (HQS-1 variable-capacitance, HQS-2 piezoresistive silicon); RTD and thermocouple interface modules available
Range
Pressure modules 0.25 in. H2O through 10,000 psi; electrical 0-20 mA or 0-30 Vdc

Field reference standard for calibrating pressure gauges, transmitters and switches; also measures current, voltage and leaks.

HQS-1 accuracy +/-0.1% to +/-0.06% of span; HQS-2 +/-0.1% to +/-0.025%.

MGP Instruments IRD 2000 (DosiCal N version) Now Mirion Technologies

dosimeter irradiator (Cs-137) for checking electronic dosimeters
Detector
Cs-137 sealed source, 370 MBq; three exposure positions per irradiation module
Detects
gamma (Cs-137)
Range
3 mSv/h at the dosimeter; 50 µSv per check

Checks and, if needed, adjusts the response of DMC 2000, SOR and DMC 3000 dosimeters automatically.

Dose rate outside the unit under 2.5 µSv/h at 1 m; about 69 dosimeters/h per module; weight up to 80 kg.

Teledyne Hastings Instruments Model 40

power supply and digital readout for flow meters
Range
Inputs: up to four Hastings flow instruments (HFM, ST, FST prefixes) and up to two HFM-D; signal output 0-5 VDC

Powers and displays up to four Hastings flow meters, with a dial-in gas conversion factor for each channel.

3-1/2 digit LCD; 115/230 VAC; 15 VDC at +/-300 mA output to the flow instruments.

No instruments match. Try a model number, a maker, or a detector type.

Specifications are summarized from manufacturer manuals and datasheets for study and identification; check the current documentation before relying on a value. Model names and trademarks belong to their owners. NukeWorker is not affiliated with any manufacturer. Know a model that belongs here, or see an error? Let us know.